Light-tight Beta Scintillator System Twin

  • Sensitivity of system with the 0.25 mm beta scintillators,
    F-18: 20 cps/(Mbq/ml)
  • Sensitivity of system with the 0.5 mm beta scintillator,
    F-18: 77 cps/(Mbq/ml)
  • Sensitivity of system with the 2 mm beta scintillator
    (surface probe) F-18: 500 cps/(MBq/ml)
  • For C-11 and O-15 the sensitivities are markedly higher.
  • Dark counts in ambient day light for 0.25 mm
    beta scintillators: 0.5 cps
  • Dark counts in ambient day light for 0.5 mm
    beta scintillators: 2 cps
  • Dark counts in ambient day light for 2 mm
    beta scintillators: 8 cps


     

Box size

  • 44 x 30 x 13 cm, rack-mountable


Weight

  • depending on model